ISO 14117:2012 specifies test methodologies for the evaluation of the electromagnetic compatibility (EMC) of active implantable cardiovascular devices that provide one or more therapies for bradycardia, tachycardia and cardiac resynchronization.
It specifies performance limits of these devices, which are subject to interactions with EM emitters operating across the EM spectrum in the two following ranges:
0 Hz ≤ f < 450 MHz;
450 MHz ≤ f ≤ 3 000 MHz.
ISO 14117:2012 also specifies requirements for the protection of these devices from EM fields encountered in a therapeutic environment and defines their required accompanying documentation, providing manufacturers of EM emitters with information about their expected level of immunity.
WITHDRAWN
ISO 14117:2012
95.99
Withdrawal of Standard
Sep 3, 2019
PUBLISHED
ISO 14117:2019